Z-1156-2009

NurseMate with Physio Module; Remote Review, Monitoring and Charting Station; USA EPMedSystems, Inc. Cooper...

Abbott Medical / initiated 2009-02-13 / Terminated / root cause: manufacturing process

St. Jude Medical began a recall of NurseMate with Physio Module; Remote Review, Monitoring and Charting Station; USA EPMedSystems, Inc. Cooper Run Executive Park, 575 Route 73 North, Unit D... on . The reason FDA records is "Installation problem: there is a risk that the monitor may not be adequately secured to the stand, resulting in the monitor falling off which may cause injury". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1156-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
NurseMate with Physio Module; Remote Review, Monitoring and Charting Station; USA EPMedSystems, Inc. Cooper Run Executive Park, 575 Route 73 North, Unit D, West Berlin, NJ 08090. Outer box labeled as Smiths Medical. The NurseMate Remote Review Station is an integrated review station for the EP WorkMate Computer recording system that allows a separate user to review and edit patient study date stored on the SP Work Mate System and monitor patient data from the EP WorkMate during a Patient study.
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
Installation problem: there is a risk that the monitor may not be adequately secured to the stand, resulting in the monitor falling off which may cause injury.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
St. Jude Medical received a recall notice dated January 16, 2009 from its supplier, Smiths Medical PM, Inc. St. Jude Medical prepared and sent their own recall letter dated February 13, 2009 to affected customers. The recall letter described the issue and possible associated risks and provided instructions to customers. A St. Jude Medical representative will contact customers to facilitate replacement of the fasteners. Questions should be directed to your St. Jude Medical sales representative.
Quantity in commerce
4 units
Distribution
Nationwide Distribution -- CA, NV, and FL.
Product codes and lots
Smith Medical Part Number: 87-1785-0000; Serial Numbers: AM08080015, AM08080016, AM08080017, AM08080018, AM08080019, AM08100004, AM08100005, AM08100006, AM08100007 and AM08100008.