Z-1081-2009

Magic3 (Intermittent Catheter), Sterile. Hydrophilic 12 French Male Length REF 53612 Rochester Medical...

Rochester Medical Corp. / initiated 2009-01-14 / Terminated / root cause: packaging

Rochester Medical Corp began a recall of Magic3 (Intermittent Catheter), Sterile. Hydrophilic 12 French Male Length REF 53612 Rochester Medical, Rochester Medical Corporation, Stewartville, MN 55976... on . The reason FDA records is "Rochester Medical observed some Magic3 and HydroSil(R) catheter packages in which the catheter was caught in the package seal during the packaging process". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1081-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Rochester Medical Corp.
Product
Magic3 (Intermittent Catheter), Sterile. Hydrophilic 12 French Male Length REF 53612 Rochester Medical, Rochester Medical Corporation, Stewartville, MN 55976. Intended for use for bladder management including urine drainage, collection and measurement.
Product code
KOD Catheter, Urological
Reason for recall
Rochester Medical observed some Magic3 and HydroSil(R) catheter packages in which the catheter was caught in the package seal during the packaging process.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Consignees were contacted by telephone or in-person visit. The communication provided the reason for the recall and a list of affected products and their lot numbers, and instruction for the removal. Distributors were given instructions to notify their customers in turn. For the customers that were not reached via telephone call or in-person, a letter, "Voluntary recall of magic product samples," was used for the recipients of product samples and sent via e-mail. The letter was dated January 14, 2009 and asked customers to destroy all of their remaining sample inventory of the affected lot numbers.
Quantity in commerce
43
Distribution
Worldwide Distribution -- USA including states of CA, FL, GA, ID, IL, LA, MD, MA, MI, MN, NJ, NM, NY, OH, PA, and WA and countries of United Kingdom.
Product codes and lots
Lot # 53606907