Z-1209-2009
KyphX Xpander Inflatable Bone Tamp, size 15/3, Catalog Number K09A, Manufactured by Kyphon Sarl, Switzerland...
Medtronic, Inc. / initiated 2009-02-04 / Terminated / root cause: sterility
Medtronic Spine LLC, formerly Kyphon Inc began a recall of KyphX Xpander Inflatable Bone Tamp, size 15/3, Catalog Number K09A, Manufactured by Kyphon Sarl, Switzerland. For use in spinal surgery on . The reason FDA records is "Product not sterile: 10 units were released for distribution without undergoing sterilization process". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1209-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- KyphX Xpander Inflatable Bone Tamp, size 15/3, Catalog Number K09A, Manufactured by Kyphon Sarl, Switzerland. For use in spinal surgery.
- Product code
- HRX Arthroscope
- Reason for recall
- Product not sterile: 10 units were released for distribution without undergoing sterilization process.
- Root cause tag
- sterility
- FDA root cause
- Process change control
- Action
- Sales representatives were contacted via e-mail on February 4, 2009, and instructed to contact hospitals and inform them of the issue and sequester affected product. They were further instructed to immediately travel to the hospitals and obtain affected product for return to Medtronic LLC. A Customer Communication Notice URGENT NOTICE: DEVICE REMOVAL dated February 9, 2009, was sent to the two consignees.
- Quantity in commerce
- 10 units affected.
- Distribution
- Product was distributed to two hospitals in OH and FL.
- Product codes and lots
- Catalog number K09A, Lot number 50000231, UPN 00858196001161.