Z-1209-2009

KyphX Xpander Inflatable Bone Tamp, size 15/3, Catalog Number K09A, Manufactured by Kyphon Sarl, Switzerland...

Medtronic, Inc. / initiated 2009-02-04 / Terminated / root cause: sterility

Medtronic Spine LLC, formerly Kyphon Inc began a recall of KyphX Xpander Inflatable Bone Tamp, size 15/3, Catalog Number K09A, Manufactured by Kyphon Sarl, Switzerland. For use in spinal surgery on . The reason FDA records is "Product not sterile: 10 units were released for distribution without undergoing sterilization process". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1209-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
KyphX Xpander Inflatable Bone Tamp, size 15/3, Catalog Number K09A, Manufactured by Kyphon Sarl, Switzerland. For use in spinal surgery.
Product code
HRX Arthroscope
Reason for recall
Product not sterile: 10 units were released for distribution without undergoing sterilization process.
Root cause tag
sterility
FDA root cause
Process change control
Action
Sales representatives were contacted via e-mail on February 4, 2009, and instructed to contact hospitals and inform them of the issue and sequester affected product. They were further instructed to immediately travel to the hospitals and obtain affected product for return to Medtronic LLC. A Customer Communication Notice URGENT NOTICE: DEVICE REMOVAL dated February 9, 2009, was sent to the two consignees.
Quantity in commerce
10 units affected.
Distribution
Product was distributed to two hospitals in OH and FL.
Product codes and lots
Catalog number K09A, Lot number 50000231, UPN 00858196001161.