BD FACS Sample Prep Assistant (SPA and SPA II) Instrument, Model Number: 334080, Product is manufactured and...
BD Biosciences, Systems & Reagents began a recall of BD FACS Sample Prep Assistant (SPA and SPA II) Instrument, Model Number: 334080, Product is manufactured and distributed by BD Biosciences, San Jose, CA The... on . The reason FDA records is "To prevent a needle stick while replacing the probe, the firm has initiated a Field Correction, and issued an improved Users Guide procedure (User Bulletin, Part Number 645050)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1620-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Product
- BD FACS Sample Prep Assistant (SPA and SPA II) Instrument, Model Number: 334080, Product is manufactured and distributed by BD Biosciences, San Jose, CA The intended use for the BD FACS Sample Prep Assistant (SPA and SPA II) is to prepare human whole blood for flow cytometry.
- Product code
- JQW Station, Pipetting And Diluting, For Clinical Use
- Reason for recall
- To prevent a needle stick while replacing the probe, the firm has initiated a Field Correction, and issued an improved Users Guide procedure (User Bulletin, Part Number 645050).
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, BD Biosciences, sent an information letter dated December 2008 with an enclosed New Probe Replacement Service/User Bulletin, to the customers. The letter described the product, problem and action to be taken. The customers were instructed to review the attached User Bulletin (Part Number 645050); ensure that their laboratory follows the updated procedure for removing and replacing probe; place the User Bulletin in their BD FACS SPA or SPA II User's Guide; always follow the instructions in User Bulletin to properly remove and replace probe, and complete and return the Effectivity Form via fax to 408-577-1409. Should you have any comments or questions, please contact BD Biosciences at 1-877-232-8995 in the US. Outside the US, contact your regional BD office or field service engineer for assistance.
- Quantity in commerce
- 70
- Distribution
- Worldwide distribution: USA and countries including: Hong Kong, Australia, Belgium, Canada, El Salvador, Hong Kong, Japan, and Mexico.
- Product codes and lots
- Lot Number: T0040