PLAC Test Calibrator Kit, Five-level liquid calibrator set for use with the diaDexus PLAC Test Reagent Kit...
DiaDexus, Inc began a recall of PLAC Test Calibrator Kit, Five-level liquid calibrator set for use with the diaDexus PLAC Test Reagent Kit, assembled in a kit box with an outer label, Catalog... on . The reason FDA records is "Premature Expiration-- the product is projected to fall short of expected shelf life, and may result in inaccurate results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1205-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diadexus, Inc.
- Product
- PLAC Test Calibrator Kit, Five-level liquid calibrator set for use with the diaDexus PLAC Test Reagent Kit, assembled in a kit box with an outer label, Catalog numbers 90108 and 10-0108, manufactured by diaDexus inc., South San Francisco, CA. Intended to establish points of reference that are used in the determination of values in the determination of Lp-PLA2 by the PLAC Test Reagent Kit.
- Product code
- NOE Test, System, Immunoassay, Lipoprotein-Associated Phospholipase A2
- Reason for recall
- Premature Expiration-- the product is projected to fall short of expected shelf life, and may result in inaccurate results.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Distributors and Direct accounts were notified by phone followed by Notification letter and certificate of destruction via FedEx. International customers were notified via e-mail with PDF copies of notices, to be followed by hard copies via FedEx. Customers are requested to record the number of kits (full and partial ) they had on the Certificate of Destruction and to dispose of that number of kits with two people signing the form, and then to fax that form to diaDexus.
- Quantity in commerce
- 105 kits
- Distribution
- Worldwide Distribution
- Product codes and lots
- lot number 048091, expiration date 2009-09.