Z-0888-2011

Varian Medical Systems Floorstand. An accessory of the Optical Guidance Platform, Model number HZ5 for use...

Varian Medical Systems, Inc. / initiated 2009-01-26 / Terminated / root cause: design

Varian Medical Systems Oncology Systems began a recall of Varian Medical Systems Floorstand. An accessory of the Optical Guidance Platform, Model number HZ5 for use with a charged particle accelerator (such as the... on . The reason FDA records is "Potential patient impact -- limit switches for Floorstand may be inadequate for some models and allow for potential impact". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0888-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
Varian Medical Systems Floorstand. An accessory of the Optical Guidance Platform, Model number HZ5 for use with a charged particle accelerator (such as the Clinac 600C).
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Potential patient impact -- limit switches for Floorstand may be inadequate for some models and allow for potential impact.
Root cause tag
design
FDA root cause
Device Design
Action
Varian sent an urgent device correction notification letter 01/26/2009 to all affected users, with a description of the problem and user corrective action steps via certified mail. The letter will also be distributed to Varian Sales, Marketing and Service organizations.
Quantity in commerce
3 units
Distribution
Product was distributed to 3 consignees in Florida and Wisconsin.
Product codes and lots
Codes HZ5071, HZ59014 and HZ59017.