Z-0555-2010

Zimmer, Insall/Burstein II Modular Knee System, Tibial Wedge and 2 Screws, Size 59, 7 Deg. x 9.5 mm Height...

Zimmer GmbH / initiated 2009-02-26 / Terminated / root cause: manufacturing process

Zimmer Inc. began a recall of Zimmer, Insall/Burstein II Modular Knee System, Tibial Wedge and 2 Screws, Size 59, 7 Deg. x 9.5 mm Height, Tivanium TI-6AL-4V Alloy, sterile; REF on . The reason FDA records is "Smaller diameter screws than required were included in the package". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0555-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer, Insall/Burstein II Modular Knee System, Tibial Wedge and 2 Screws, Size 59, 7 Deg. x 9.5 mm Height, Tivanium TI-6AL-4V Alloy, sterile; REF 5220-62-01.
Product code
HSX Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer
Reason for recall
Smaller diameter screws than required were included in the package.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Zimmer notified consignees by recall letter dated 2/25/09 instructing them to cease use of the device and to return it. Zimmer distributors and international accounts were notified by email on 2/26/09.
Quantity in commerce
13
Distribution
USA (Connecticut, Indiana, Pennsylvania), Australia, Serbia, Spain, Switzerland and the United Kingdom.
Product codes and lots
Lot 07878240.