Z-1122-2009

ILED Surgical Light, Model numbers: ILED3, ILED5 and ILED5K

Hill-Rom Inc. / initiated 2009-02-27 / Terminated

Trumpf Medical Systems, Inc. began a recall of ILED Surgical Light, Model numbers: ILED3, ILED5 and ILED5K on . The reason FDA records is "Cracks may form around the screw connections of the cover plates causing brittleness, which may lead to fragments breaking off and falling into the operating area". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1122-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Inc.
Product
ILED Surgical Light, Model numbers: ILED3, ILED5 and ILED5K
Product code
FSY Light, Surgical, Ceiling Mounted
Reason for recall
Cracks may form around the screw connections of the cover plates causing brittleness, which may lead to fragments breaking off and falling into the operating area.
FDA root cause
Other
Action
Consignees were notified via an Urgent Safety Information letter starting 02/27/2009. The Recall Notice asks customers to visually inspect the ILED covers for cracks and in particular to focus on the corners. If cracking is visible they are asked to remedy by taping with a clear tape over the cracks and report it to TRUMPF Medical Systems, Inc. immediately. TRUMPF will be replacing all covers with a newly designed and tested cover. Original priority in the United States will be set to start with the oldest units affected. If TRUMPF receives notification that cracking does exist then that facility will rise to the highest priority. Additionally, TRUMPF will issue a new approved cleaning and disinfection instruction, since investigations have shown that the cracking is significantly influenced by the use of certain disinfectants containing alcohol. The operator manual previously stated to avoid using alcohol-based cleaners on some parts of the ILED, but after reviewing this it was determined that it was not clearly stated to avoid use on the covers.
Quantity in commerce
4480 units (540 in U.S.)
Distribution
Worldwide Distribution.
Product codes and lots
Serial numbers: 100268621 to 100717475