Z-1184-2009

Stryker PainPump2, 250 mL PainPump with 2.5 inch ExFen Catheter and Luer Lock Tubing Set, 5 per box, sterile...

Stryker Corp. / initiated 2009-02-11 / Terminated / root cause: manufacturing process

Stryker Instruments Div. of Stryker Corporation began a recall of Stryker PainPump2, 250 mL PainPump with 2.5 inch ExFen Catheter and Luer Lock Tubing Set, 5 per box, sterile, Stryker, Kalamazoo, MI; REF 525-158. The Stryker... on . The reason FDA records is "An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1184-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker PainPump2, 250 mL PainPump with 2.5 inch ExFen Catheter and Luer Lock Tubing Set, 5 per box, sterile, Stryker, Kalamazoo, MI; REF 525-158. The Stryker PainPump2 is an electromechanical pump designed to deliver a controlled amount of medication to the patient for pain management.
Product code
FRN Pump, Infusion
Reason for recall
An electronic failure during preparation of the unit, or during patient use, may cause the pump to shut down.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
A "Medical Device Recall Notification" letter dated February 11, 2009 was issued to consignees via phone, mail and/or email. The letter described the issue and instructed customers to discontinue use of the recalled product, destroy it according to instructions provided with the notification and return the "Business Reply Form" by fax to Stryker Instruments Regulatory Department, 1-866-521-2762. Further questions should be addressed to Stryker Instruments at 1-800-800-4236 extension 4086.
Quantity in commerce
24,675 pumps
Distribution
Worldwide Distribution -- USA including states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, WA, WI, WV and WY and countries of Argentina, Australia, Brazil, Canada, England, Singapore and South Africa.
Product codes and lots
Lot Number: 08316012.