BD FACSCount Controls Kit, An Automated Differential Cell Counter, Model/Catalog Number: P/N: 340166, Product...
BD Biosciences, Systems & Reagents began a recall of BD FACSCount Controls Kit, An Automated Differential Cell Counter, Model/Catalog Number: P/N: 340166, Product is manufactured and distributed by BD... on . The reason FDA records is "The product is mislabeled, in that, zero/low controls were mislabeled as medium/high, and the medium/high controls were labeled zero/low". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1007-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Product
- BD FACSCount Controls Kit, An Automated Differential Cell Counter, Model/Catalog Number: P/N: 340166, Product is manufactured and distributed by BD Biosciences, San Jose, CA The reagents are for in vitro diagnostic use on a BD FACSCount instrument.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- The product is mislabeled, in that, zero/low controls were mislabeled as medium/high, and the medium/high controls were labeled zero/low.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- On 1/29/09, the firm initiated the recall and its notification was via Important Product Information letters identifying the affected product and explaining the reason for the recall. Customers who experienced error codes with the affected product were to contact their local BD Biosciences representative or distributor to receive replacement BD FACSCount controls.
- Quantity in commerce
- 17 units
- Distribution
- Nationwide Distribution -- Including states of WA, MO, NV, CA, NY, PA, AL, and OR.
- Product codes and lots
- Lot Number: 89031721