Z-1016-2010
Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System; Model Number/Catalog Number...
Biogenex Laboratories, Inc. / initiated 2009-01-23 / Terminated / root cause: design
Biogenex Laboratories began a recall of Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System; Model Number/Catalog Number: AS4010B; The product is manufactured by: BioGenex... on . The reason FDA records is "Under certain conditions, product system failure has the potential to cause operator injury due to splash of Hemo-De solution". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1016-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biogenex Laboratories, Inc.
- Product
- Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System; Model Number/Catalog Number: AS4010B; The product is manufactured by: BioGenex Laboratories, Inc. 4600 Norris Canyon Road San Ramon, CA 94583 and Distributed by: Abbott Molecular, 1300 E. Touhy Ave - 2E, Des Plaines, IL 60018 Xmatrx Automated Staining System is intended for in vitro diagnostic use. It is suitable for fluorescent hybridization (FISH) applications.
- Product code
- KEY Stainer, Tissue, Automated
- Reason for recall
- Under certain conditions, product system failure has the potential to cause operator injury due to splash of Hemo-De solution.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 1/23/09, the firm issued an Advisory Notice to all its direct consignees, informing them of the affected product and providing instructions on the recall. Please contact BioGenex Technical Support at 1-800-421-4149 for any assistance you may need.
- Quantity in commerce
- 12 units
- Distribution
- The firm distributed the product to one (1) distributor as a consignee of our product. The firm indicated that no customer has received the product directly from BioGenex, due to the fact that the product is exclusive distributed by Abbott Molecular.
- Product codes and lots
- Serial Numbers: AS40211, AS40226, AS40230, AS40231, AS40232, AS40233, AS40234, AS40235, AS40236, and AS40237