ALM Prismatic Surgical Light 9551 DF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide...

Maquet Cardiovascular Us Sales, LLC / initiated 2009-02-23 / Terminated / root cause: design

MAQUET Inc. began a recall of ALM Prismatic Surgical Light 9551 DF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide visible illumination for the surgical field or for the... on . The reason FDA records is "The screws holding the Prismatic triple surgical lights may break due to dynamic multidirectional bending stress. The firm has received one complaint report". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1153-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
ALM Prismatic Surgical Light 9551 DF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide visible illumination for the surgical field or for the examination of the patient. PRC is designed to eliminate shadow, reduce reflected heat, and illuminate surgical incisions of all depths during surgical procedures.
Product code
FTD Lamp, Surgical
Reason for recall
The screws holding the Prismatic triple surgical lights may break due to dynamic multidirectional bending stress. The firm has received one complaint report.
Root cause tag
design
FDA root cause
Device Design
Action
An "Urgent Device Correction" letter dated February 23, 2009 was mailed by FedEx Tracking to all affected consignees. The letter described the issue, potential hazard and recommended precaution. Customers were instructed to complete the enclosed "Customer Response Fax Form" and return it via fax (1-732-667-1903) to Maquet, Inc. Maquet, Inc. will schedule repair service of the affected units once the "Customer Response Fax Form" is received. Direct questions to your local Maquet, Inc. Representative.
Quantity in commerce
54 of all units, both article numbers
Distribution
Nationwide Distribution.
Product codes and lots
Article Number: 567014999; Serial Numbers: 1104, 1105, 1106, 25163, 25340, 25341, 25890, 26012, 26417, 26418, 26686, 27251, 27753, 27854, 27855, 28096, 28097, 28813, 29652, 29950, 30087, 30152, 30333-30334, 30739, 31250, 31381, 32333, 32334, 32336, 32580, 32708, 32709 and 32800.