Alm Prismatic Surgical Light 5004 SF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide...
MAQUET Inc. began a recall of Alm Prismatic Surgical Light 5004 SF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide visible illumination for the surgical field or for the... on . The reason FDA records is "The screws holding the Prismatic triple surgical lights may break due to dynamic multidirectional bending stress. The firm has received one complaint report". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1154-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Alm Prismatic Surgical Light 5004 SF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide visible illumination for the surgical field or for the examination of the patient. PRC is designed to eliminate shadow, reduce reflected heat, and illuminate surgical incisions of all depths during surgical procedures.
- Product code
- FTD Lamp, Surgical
- Reason for recall
- The screws holding the Prismatic triple surgical lights may break due to dynamic multidirectional bending stress. The firm has received one complaint report.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An "Urgent Device Correction" letter dated February 23, 2009 was mailed by FedEx Tracking to all affected consignees. The letter described the issue, potential hazard and recommended precaution. Customers were instructed to complete the enclosed "Customer Response Fax Form" and return it via fax (1-732-667-1903) to Maquet, Inc. Maquet, Inc. will schedule repair service of the affected units once the "Customer Response Fax Form" is received. Direct questions to your local Maquet, Inc. Representative.
- Quantity in commerce
- 54 of all units, both article numbers
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Article Number: 567030999; Serial Numbers: 6041, 6042, 6043, 6044, 6045, 6046, 6047, 6048, 6049, 6050, 6051, 6052, 6053, 6054, 6057, 6058, 6059, 6060, 6061 and 6062.