GE Healthcare Signa Ovation 0.35T, Model Numbers: Ovation 1-3 (2276937), Ovation 4 (2377062-2, 2377062-5...
GE Medical Systems, LLC began a recall of GE Healthcare Signa Ovation 0.35T, Model Numbers: Ovation 1-3 (2276937), Ovation 4 (2377062-2, 2377062-5, 2377062-8, or 5118172), and Ovation 5 (5148725). The... on . The reason FDA records is "GE Healthcare has identified a potential pinch point hazard on Ovation MR Scanner when using CTL Array XL Body Flex and Opened Body coil. A patient's hand may be pinched between the magnet bore ceiling cover and the CTL...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1722-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Signa Ovation 0.35T, Model Numbers: Ovation 1-3 (2276937), Ovation 4 (2377062-2, 2377062-5, 2377062-8, or 5118172), and Ovation 5 (5148725). The 0.35T Signa Ovation with Excite Magnetic Resonance system is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The 0.35T Signa Ovation with Excite Magnetic Resonance system is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the body including, but not limited to, the musculoskeletal, vascular, cardiac, and neuron systems.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- GE Healthcare has identified a potential pinch point hazard on Ovation MR Scanner when using CTL Array XL Body Flex and Opened Body coil. A patient's hand may be pinched between the magnet bore ceiling cover and the CTL Array, XL Body Flex, or Open Body coil if the patient puts his/her hand on the top of the coil. 2/20/09: Recall being expanded to include the OpenSpeed Systems.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- GE Healthcare issued a "Product Safety Notification" letter dated October 23, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology/Cardiology & Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information for the Signa Ovation 0.35T Systems. A second GE Healthcare "Urgent Medical Device Correction" letter dated March 5, 2009 was sent 3/9/09 to consignees which addressed the expanded recall to the Signa OpenSpeed0.7T MR Systems. For further questions, contact your local GE Healthcare sales or service representative.
- Quantity in commerce
- 308 (143 US / 142 OUS / 23 TBD)
- Distribution
- Worldwide Distribution-- US (including states of AL, ALZ, AR, CA, CO,CT, DE, FL, FA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MT, MO, NE,NH, NV, NJ, NY, NC, OH, OK, OR, PA, PR, SC SD, TN, TX, VA, WA, WV and WI), CHINA, CYPRUS, ARGENTINA, BRAZIL EGYPT, DENMARK, FEDERAL REPUBLIC OF GERMANY, FEDERAL REPUBLIC OF FRANCE, GET 7 D CHINA, GREECE, HUNGARY, ITALY, IVORY COAST, JAPAN, KOREA, KUWAIT, MALAWI, MEXICO, OMAN, PANAMA, PHILIPPINES, RUSSIAN FEDERATION, SAUDI ARABIA, SPAIN , SWITZERLAND, TURKEY, UKRAINE, UNITED ARAB EMERITUS, UNITED KINGDOM and VENEZUELA.
- Product codes and lots
- Model Number: 2276937; Serial Numbers: 00000869965YM4, 00000801207YM2, 00000813952YM9, 00000952559YM3, 00000236062MR6, 00000869966YM2, 00000801206YM4, 00000813945YM3, 00000813950YM3, 00000869958YM9, 00000952561YM9, 00000813958YM6, 00000919518YM1, 00000813944YM6, 00000837353YM2, 00000813960YM2, 00000837350YM8, 00000264247MR1, 00000897708YM4, 00000897716YM7, 00000919512YM4, 00000919526YM4, 00000026540VE5, 00000919509YM0, 00000813959YM4, 00000801204YM9, 00000813947YM9, 00000952560YM1, 00000952565YM0, 00000897711YM8, 00000897717YM5, 00000919515YM7, 00000952552YM8, 00000813951YM1, 00000897701YM9, 00000869949YM8, 00000837359YM9, 00000919519YM9, 00000813946YM1, 00000869956YM3, 00000897706YM8, 00000952563YM5, 00000869947YM2, 00000801199YM1, 00000801193YM4, 00000837362YM3, 00000952548YM6, 00000952549YM4, 00000837360YM7, 00000952547YM8, 00000837346YM6, 00000869952YM2, 00000869954YM8, 00000919514YMO, 00000919524YM9, 00000837358YM1, 00000919521YM5, 00000869962YM1, 00000919513YM2, 00000897705YM0, 00000952555YM1, 00000813943YM8, 00000869961YM3, 00000813956YM0, 00000801203YM1, 00000801208YM0, 00000801209YM8, 00000813948YM7, 00000813953YM7, 00000813954YM5, 00000837347YM4, 00000837351YM6, 00000837352YM4, 00000837354YM0, 00000837357YM3, 00000837361YM5, 00000837363YM1, 00000837364YM9, 00000837365YM6, 00000869950YM6, 00000869951YM4, 00000869953YM0, 00000869955YM5, 00000869957YM1, 00000869960YM5, 00000869963YM9, 00000897698YM7, 00000897699YM5, 00000897700YM1, 00000897702YM7, 00000897703YM5, 00000897704YM3, 00000897707YM6, 00000897715YM9, 00000919507YM4, 00000919510YM8, 00000919516YM5, 00000919517YM3, 00000919520YM7, 00000919523YM1, 00000952550YM2, 00000952553YM6, 00000952557YM7, 00000952558YM5, 00000952562YM7, 00000952566YM8, 00000952554YM4, 00000919511YM6, 00000869948YM0, 00000897714YM2, 00000813957YM8, 00000813955YM2, 00000897712YM6, 00000813949YM5; Model Number: 5118172; Serial Numbers; 00000000564YR5, 00000000565YR2, 00000000563YR7, 00000000561YR1; Model Number: 2377062-2; Serial Number: 00000963007YM0; Model Number: 2377062-5; Serial Numbers: 00000000120YR6, 00000000123YR0, 00000000132YR1, 00000000124YR8, 00000000137YR0, 00000000130YR5, 00000000134YR7, 00000000138YR8, 00000962592YM2, 00000000146YR1, 00000000117YR2, 00000000116YR4, 00000000118YR0, 00000943610YM6, 00000000152YR9, 00000000140YR4, 00000000110YR7, 00000000135YR4, 00000000106YR5, 00000000129YR7, 00000000141YR2, 00000000125YR5, 00000000142YR0, 00000000113YR1, 00000000150YR3, 00000000105YR7, 00000000131YR3, 00000000139YR6, 00000000155YR2, 00000000108YR1, 00000000143YR8, 00000000101YR6, 00000000109YR9, 00000000153YR7, 00000951021YM5, 00000000119YR8, 00000000136YR2, 00000000128YR9, 00000000100YR8, 00000000112YR3, 00000000121YR4, 00000000154YR5; Model Number: 2377062-8; Serial Numbers: 00000000776YR5, 00000000792YR2, 00000000793YR0, 00000000801YR1, 00000000803YR7, 00000000829YR2, 00000000836YR7, 00000000758YR3, 00000000765YR8, 00000000755YR9, 00000000756YR7, 00000000774YR0, 00000000784YR9, 00000000785YR6, 00000000798YR9, 00000000827YR6, 00000000816YR9, 00000000779YR9, 00000000828YR4, 00000000789YR8, 00000000838YR3, 00000000773YR2, 00000000788YR0, 00000000796YR3, 00000000763YR3, 00000000766YR6, 00000000767YR4, 00000000787YR2, 00000000790YR6, 00000000800YR3, 00000000834YR2, 00000000840YR9, 00000000830YR0, 00000000832YR6, 00000000786YR4, 00000000764YR1, 00000000775YR7, 00000000818YR5, 00000000768YR2, 00000000757YR5, 00000000760YR9, 00000000762YR5, 00000000783YR1, 00000000809YR4, 00000000782YR3, 00000000812YR8, 00000000826YR8, 00000000835YR9, 00000000770YR8, 00000000771YR6, 00000000825YR0, 00000000759YR1, 00000000821YR9, 00000000841YR7, 00000000831YR8, 00000000808YR6, 00000000823YR5, 00000000824YR3, 00000000772YR4, 00000000810YR2, 00000000837YR5, 00000000769YR0, 00000000815YR1, 00000000833YR4, 00000000804YR5, 00000000778YR1, 00000000780YR7, 00000000795YR5, 00000000799YR7, 00000000805YR2, 00000000807YR8, 00000000819YR3, 00000000839YR1, 00000000811YR0, 00000000761YR7, 00000000777YR3, 00000000781YR5, 00000000791YR4, 00000000797YR1, 00000000794YR8; and Model Number: 5148725; Serial Numbers: 00000154413HM9, 00000176551HM0, 00000139846HM0, 00000192389HM5, 00000177143HM5, 00000193109HM6, 00000148905HM3, 00000147526HM8, 00000138166HM4, 00000000SPI275, 00000188489HM9, 00000167482HM9, 00000169104HM7, 00000171692HM7, 00000157419HM3, 00000266430MR8, 00000155516HM8, 00000155515HM0, 00000160809HM0, 00000175023HM1, 00000150718HM5, 00000162563HM1, 00000164072HM1, 00000165772HM5, 00000172421HM0, 00000172816HM1, 00000172869HM0, 00000173908HM5, 00000177792HM9, 00000178519HM5, 00000179361HM1, 00000180127HM3, 00000180746HM0, 00000182071HM1, 00000182249HM3, 00000182597HM5, 00000183130HM4, 00000187365HM2, 00000188810HM6, 00000189105HM0, 00000191017HM3, 00000191154HM4, 00000196647HM2, 00000200577HM5, 00000265078MR6, 00000157418HM5, 00000174348HM3, 00000189943HM4, 00000181736HM0, 00000146891HM7, 00000158641HM1, 00000147906HM2, 00000169833HM1, 00000170940HM1, 00000190359HM0, 00000191544HM6, 00000193220HM1, 00000193936HM2, 00000195772HM9, 00000144797HM8, 00000150152HM7, 00000170109HM3, 00000181735HM2, 00000197645HM5, 00000168323HM4 and 00000181517HM4.