Z-1235-2009

syngo Imaging, picture archiving and communications system (PACS), Model number 10 014 063

Siemens Medical Solutions USA, Inc. / initiated 2009-01-12 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of syngo Imaging, picture archiving and communications system (PACS), Model number 10 014 063 on . The reason FDA records is "Incorrect and/or incomplete data: Several potential safety issues while working with the syngo Image product including, but not limited to, data loss, errors in patient data due to different patient IDs, delays in image...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1235-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
syngo Imaging, picture archiving and communications system (PACS), Model number 10 014 063
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Incorrect and/or incomplete data: Several potential safety issues while working with the syngo Image product including, but not limited to, data loss, errors in patient data due to different patient IDs, delays in image availability, inaccurate results during image measurements, incorrect review or diagnosis of displayed or printed images, measurement inaccuracies from the host modality and image
Root cause tag
software
FDA root cause
Software design
Action
syngo Imaging has issued an Addendum to the User Manual, which was distributed to affected customers via Update Instructions IM118/08/S in November, 2008. Two safety related hotfixes (VB20G_HF4 and VB30A_HF26) will be released to those affected only after Update Instructions IM118/08/S has been performed. The addendum informs the user about safety relevant notes which pertain to syngo Imaging versions VB20G and VB30A. Update is a prerequisite for the following safety related hotfixes: VB20G_HF4 and VB30A_HF26.
Quantity in commerce
55 units
Distribution
Nationwide Distribution.
Product codes and lots
Model number 10 014 063. Serial numbers: 1005, 1039, 1040, 1041, 1043, 1049, 1053, 1055, 1056, 1057, 1058, 1061, 1064, 1065, 1067, 1071, 1074, 1076, 1079, 1080, 1081, 1082, 1083, 1084, 1086, 1090, 1105, 1107, 1112, 1114, 1116, 1119, 1124, 1174, 1188, 1189, 1202, 1203, 1210, 1218, 1226, 1227, 1229, 1240, 1246, 1247, 1258, 1269, 1271, 1278, 1279, 1310, 1316, 1329, and 1334.