Screwdriver for 3.5mm Screws, DARCO system, instrument for use in orthopaedic implant surgery. REF: DC4261...
Wright Medical Technology Inc began a recall of Screwdriver for 3.5mm Screws, DARCO system, instrument for use in orthopaedic implant surgery. REF: DC4261, Non-sterile, Rx only, 1 each, Wright Medical... on . The reason FDA records is "The package insert lists 3 sterilization procedures, 2 of which are not validated. Pre-vacuum steam sterilization is validated". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1208-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Wright Medical Technology Inc
- Product
- Screwdriver for 3.5mm Screws, DARCO system, instrument for use in orthopaedic implant surgery. REF: DC4261, Non-sterile, Rx only, 1 each, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The package insert lists 3 sterilization procedures, 2 of which are not validated. Pre-vacuum steam sterilization is validated.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Wright Medical notified their distributors of the recall on 03/11/2009 by letter explaining the problem and requesting return of the product from their inventory and from their consignees. The letter included a "fax back" response page to ensure the receipt of the notice. The hospitals and surgeons were notified of the recall also on 03/11/2009. All letters were delivered by Fed Ex.
- Quantity in commerce
- 682 devices
- Distribution
- Worldwide distribution: USA, Canada, Germany, France, Belgium, the Netherlands, Italy, Finland, Austria, Denmark, Norway, Turkey, South Africa, United Kingdom, and Australia.
- Product codes and lots
- Lot numbers: 019598209, 038504328, 058486628, 058507733, 058516752, 067422530, 088538652, 098538652, 107439501, 107442247, 117442247, 0712180, S0701050