Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1263-2009

Carina Home Ventilator. The device is used as a continuous respirator/ventilator for home use

Draeger, Inc. / initiated 2009-01-21 / Terminated / root cause: software

Draeger Medical, Inc. began a recall of Carina Home Ventilator. The device is used as a continuous respirator/ventilator for home use on . The reason FDA records is "The device may not generate a pilot line disconnect alarm if the pilot line becomes disconnected, which may potentially result in a substantial reduction in tidal volume delivery". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1263-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Carina Home Ventilator. The device is used as a continuous respirator/ventilator for home use.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
The device may not generate a pilot line disconnect alarm if the pilot line becomes disconnected, which may potentially result in a substantial reduction in tidal volume delivery.
Root cause tag
software
FDA root cause
Software design
Action
An "Urgent - Medical Device Recall" Letter dated January 2009 was issued to consignees via certified mail. The letter described the issue, specific conditions, background information, and precautions for customers. An additional letter dated January 22, 2009 was included with the "Urgent - Medical Device Recall" letter which instructed consignees to inform an enduser of the contents of the notification letter because some of the consignees are home care device/service providers. When the new hardware/software solution is available, affected devices will be updated free of charge. Direct questions about the notification letter to Drager Medical, Inc. by calling 1-800-543-5047 (press "1" at the prompt and then press "2349#"). Direct questions regarding updating your device when the new software is available, contact Drager Service Technical Support at 1-800-543-5047 (press "4" at the prompt).
Quantity in commerce
40 units
Distribution
Nationwide Distribution -- states of CA, FL, GA, HI, IL, LA, MA, MN, OH, and TX.
Product codes and lots
Catalog Numbers 5704757 and 5704758.