Z-1318-2009

Mammo Test Model number 10144185, x-ray guided stereotactic biopsy system

Siemens Medical Solutions USA, Inc. / initiated 2009-01-21 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Mammo Test Model number 10144185, x-ray guided stereotactic biopsy system on . The reason FDA records is "Table may unintentionally lift during procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1318-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Mammo Test Model number 10144185, x-ray guided stereotactic biopsy system
Product code
JAQ System, Applicator, Radionuclide, Remote-Controlled
Reason for recall
Table may unintentionally lift during procedure
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens Medical Solutions issued a Safety Advisory Notice dated 1/22/09 which informs the customers of the potential issues and recommends that they stop using their MammoTest System until the prone table has been repaired. Linear Medical visited the customers to repair the weld joint on the prone table by January 23, 2009.
Quantity in commerce
4 units
Distribution
USA distribution to IL, NY, SC, and WA
Product codes and lots
Serial numbers 10102, 10103, 10105, and 10106