Z-1252-2009

Servo Ventilator 300 (Model SV300) and Servo Ventilator 300 (Model SV300A). The SV300 is intended for general...

Maquet Cardiovascular Us Sales, LLC / initiated 2009-03-23 / Terminated / root cause: design

MAQUET Inc. began a recall of Servo Ventilator 300 (Model SV300) and Servo Ventilator 300 (Model SV300A). The SV300 is intended for general and critical ventilatory care for use on adults... on . The reason FDA records is "The expiratory valve in the SV 300 ventilator may stay in the open position if the printed circuit board fails". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1252-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Servo Ventilator 300 (Model SV300) and Servo Ventilator 300 (Model SV300A). The SV300 is intended for general and critical ventilatory care for use on adults, pediatric, infant and neonatal patients. The unit is designed to be used at the bedside for intra-hospital transport.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
The expiratory valve in the SV 300 ventilator may stay in the open position if the printed circuit board fails.
Root cause tag
design
FDA root cause
Device Design
Action
An "Urgent: Device Correction" letter dated March 23, 2009 was sent by Federal Express to customers. The letter described the affected product, problem, indications, potential hazard and actions for users. Customers were instructed to complete the "Device Correction -- Customer Fax Response" form and return via fax (732-667-1903) to Maquet Inc. RA/QA. Direct questions to your local Maquet Inc. representative.
Quantity in commerce
7,847 units
Distribution
Nationwide Distribution.
Product codes and lots
All units manufactured prior to June 2002.