Gendex VixWin Platinum Imaging Software, Versions 1.0, 1.1 and 1.2; Gendex Dental Systems, 901 West Oakton...
Gendex Dental Systems began a recall of Gendex VixWin Platinum Imaging Software, Versions 1.0, 1.1 and 1.2; Gendex Dental Systems, 901 West Oakton Street, Des Plaines, IL 60018. Intended for use of... on . The reason FDA records is "There is a compatibility issue when the VixWin Platinum versions 1.0/1.1/1.2 software is used in conjunction with Easy Dental, PracticeWorks, EagleSoft and SoftDent practice management software, resulting in patient...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1548-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Gendex Corp.
- Product
- Gendex VixWin Platinum Imaging Software, Versions 1.0, 1.1 and 1.2; Gendex Dental Systems, 901 West Oakton Street, Des Plaines, IL 60018. Intended for use of the software is to control capture, display, treatments, analysis and saving of x-ray digital images.
- Product code
- MUH System, X-Ray, Extraoral Source, Digital
- Reason for recall
- There is a compatibility issue when the VixWin Platinum versions 1.0/1.1/1.2 software is used in conjunction with Easy Dental, PracticeWorks, EagleSoft and SoftDent practice management software, resulting in patient files that may contain images that are mixed between patients.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Gendex issued an "Urgent Medical Device Report" dated March 6, 2009 to the direct account dealers of the affected device via first class mail. The noticed described the issue and provided a software update directed to fixing the problem. In addition, customers were asked to complete and return an Acknowledgement Form by fax to 1-847-718-0601. Further questions should be directed to the Gendex Technical Support team at 1-888-275-5286.
- Quantity in commerce
- 1,256 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Software versions 1.0, 1.1 and 1.2.