Z-1239-2009

Edwards Lifesciences Swan-Ganz Intellicath CCO/VIP Thermodilution Catheter with AMC Thromboshield (an...

Edwards Lifesciences LLC / initiated 2009-02-09 / Terminated / root cause: software

Edwards Lifesciences, Llc began a recall of Edwards Lifesciences Swan-Ganz Intellicath CCO/VIP Thermodilution Catheter with AMC Thromboshield (an Antimicrobial *Heparin Coating) For use with Vigilance... on . The reason FDA records is "Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode"". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1239-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
Edwards Lifesciences Swan-Ganz Intellicath CCO/VIP Thermodilution Catheter with AMC Thromboshield (an Antimicrobial *Heparin Coating) For use with Vigilance Monitor. REF: 139HF75 For use inpatients who require hemodynamic monitoring. Intended to be used in combination with clinical pressure monitoring equipment to measure right heart and pulmonary artery pressures, and with the Vigilance Monitor to measure cardiac output.
Product code
KRA Catheter, Continuous Flush
Reason for recall
Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".
Root cause tag
software
FDA root cause
Software design
Action
Edwards initiated the worldwide recall via Urgent Customer Notification letters dated February 9, 2009 and are being sent by Federal Express starting on February 10, 2009. These letters are requesting that customers cease using Swan-Ganz" Continuous Cardiac Output (CCO) Thermodilution Catheter products from the indicated lots and that they return all unused product. Edwards Lifesciences requested that the customer provide written verification of stock and identify any unused product to be returned. Questions should be directed to Technical Service at 800-822-9837.
Quantity in commerce
107 units
Distribution
Worldwide Distribution -- AL, AR, AZ, CA, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Hawaii, and Puerto Rico and countries of Canada, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, UK, Greece, Ireland, Israel, Italy, Kuwait, Netherlands, Norway, Portugal, Sweden, Colombia, Australia, and Singapore.
Product codes and lots
Lot 58619322 - expiration date between May 2010 and November 2010.