Edwards Lifesciences Swan-Ganz CCOombo CCO/SvO2/VIP Thermodilution Catheter For use with Vigilance Monitor...
Edwards Lifesciences, Llc began a recall of Edwards Lifesciences Swan-Ganz CCOombo CCO/SvO2/VIP Thermodilution Catheter For use with Vigilance Monitor. This package contains: One catheter and one... on . The reason FDA records is "Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode"". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1242-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Edwards Lifesciences LLC
- Product
- Edwards Lifesciences Swan-Ganz CCOombo CCO/SvO2/VIP Thermodilution Catheter For use with Vigilance Monitor. This package contains: One catheter and one volume-limited syringe. REF: 746F8 For use inpatients who require hemodynamic monitoring. Intended to be used in combination with clinical pressure monitoring equipment to measure right heart and pulmonary artery pressures, and with the Vigilance Monitor to measure cardiac output.
- Product code
- DYG Catheter, Flow Directed
- Reason for recall
- Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Edwards initiated the worldwide recall via Urgent Customer Notification letters dated February 9, 2009 and are being sent by Federal Express starting on February 10, 2009. These letters are requesting that customers cease using Swan-Ganz" Continuous Cardiac Output (CCO) Thermodilution Catheter products from the indicated lots and that they return all unused product. Edwards Lifesciences requested that the customer provide written verification of stock and identify any unused product to be returned. Questions should be directed to Technical Service at 800-822-9837.
- Quantity in commerce
- 37 units
- Distribution
- Worldwide Distribution -- AL, AR, AZ, CA, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Hawaii, and Puerto Rico and countries of Canada, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, UK, Greece, Ireland, Israel, Italy, Kuwait, Netherlands, Norway, Portugal, Sweden, Colombia, Australia, and Singapore.
- Product codes and lots
- Lot 58617326