Z-1228-2009
MEDRAD Sterile Disposable Syringe Kit, catalog number DSK 130-Q, Angiographic Syringe for delivery of...
Medrad, Inc. / initiated 2009-01-29 / Terminated / root cause: packaging
Medrad Inc began a recall of MEDRAD Sterile Disposable Syringe Kit, catalog number DSK 130-Q, Angiographic Syringe for delivery of contrast media on . The reason FDA records is "Possible compromised sterility due to holes in styrene tray". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1228-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medrad, Inc.
- Product
- MEDRAD Sterile Disposable Syringe Kit, catalog number DSK 130-Q, Angiographic Syringe for delivery of contrast media.
- Product code
- DXT Injector And Syringe, Angiographic
- Reason for recall
- Possible compromised sterility due to holes in styrene tray
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- On 1/29/09, Medrad issued an Urgent Medical Device Recall letter. Customers were requested to check their inventory for any of the affected batch. If the customer has affected product, they are instructed to contact the firm to make arrangement for the return of the product. Non responders were sent additional notifications via FedEx on 3/2/09 and 4/1/09.
- Quantity in commerce
- 2958 kits
- Distribution
- Nationwide: AL, AZ, CA, FL, IL, LA, MA, MT, NC, NY, OH, TX, and WI.
- Product codes and lots
- catalog number DSK 130-Q. Lot number 72372