Z-1900-2011

Siemens brand ACUSON S2000 Ultrasound System with software versions: VA10, VA10A, VA10B, VA10C, VA15; Model...

Siemens Medical Solutions USA, Inc. / initiated 2009-01-14 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc. began a recall of Siemens brand ACUSON S2000 Ultrasound System with software versions: VA10, VA10A, VA10B, VA10C, VA15; Model Number: 10041461; Product is manufactured and... on . The reason FDA records is "The product has a software problem in which previous patient measurement data gets associated with another patient's image". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1900-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens brand ACUSON S2000 Ultrasound System with software versions: VA10, VA10A, VA10B, VA10C, VA15; Model Number: 10041461; Product is manufactured and distributed by Siemens Medical Solutions USA, Inc. Ultrasound Business Unit, Mountain View, CA The ACUSON SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal cardiac, Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications.
Product code
GKZ Counter, Differential Cell
Reason for recall
The product has a software problem in which previous patient measurement data gets associated with another patient's image.
Root cause tag
software
FDA root cause
Software change control
Action
The firm, SIEMENS, issued a "CUSTOMER SAFETY ADVISORY NOTIFICATION" letter to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to ensure that the first image in Exam Review is from the current patient; ensure that the Heart Cycle measurement tool is not selected on the Cale menu when entering Exam Review; be aware of this issue as described in the letter and perform four (4) or less measurements of the same measurement label. Should you have any questions, please call (650) 694-5398.
Distribution
Worldwide distribution.
Product codes and lots
All units of this model.