VARIANT II TURBO Hemoglobin Testing System with CDM 3.6T Software Version 3.5; Model Number: 270-2600...
Bio-Rad Laboratories Inc began a recall of VARIANT II TURBO Hemoglobin Testing System with CDM 3.6T Software Version 3.5; Model Number: 270-2600; Product is manufactured and distributed by Bio-Rad... on . The reason FDA records is "Sample test results for the first run appears to be correct but all subsequent results during that run were assigned incorrect barcode and tube position". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1902-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bio-Rad Laboratories Inc.
- Product
- VARIANT II TURBO Hemoglobin Testing System with CDM 3.6T Software Version 3.5; Model Number: 270-2600; Product is manufactured and distributed by Bio-Rad Laboratories, Diagnostics Group, Hercules, CA. Provides an integrated method for sample preparation, separation and the percent determination of percent hemoglobin A1c in EDTA human whole blood.
- Product code
- LCP Assay, Glycosylated Hemoglobin
- Reason for recall
- Sample test results for the first run appears to be correct but all subsequent results during that run were assigned incorrect barcode and tube position.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Bio-Rad Laboratories, Inc. issued and "Urgent: Medical Device Correction" dated January 22, 2009 to all affected consignees by FED EX, followed up by telephone calls for those not responding. Consignees were instructed on proper steps to take in dealing with the affected product. Further communication will follow to provide a solution. For further information, contact Bio-Rad Laboratories at 1-510-724-7000.
- Distribution
- Worldwide Distribution -- United States, Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Czech Republic, France, Germany, Hong Kong, Hungary, India, Italy, Mexico, New Zealand, Poland, Singapore, South Korea, Sweden, Taiwan, Thailand and the United Kingdom.
- Product codes and lots
- All units of these models.