Siemens LANTIS 6.1 Oncology Information System with MEDOne Module, Linear Medical Accelerators; Product...
Siemens Medical Solutions USA, Inc began a recall of Siemens LANTIS 6.1 Oncology Information System with MEDOne Module, Linear Medical Accelerators; Product Numbers: 4503178, 5835991, 5851055; Product is... on . The reason FDA records is "The product has a software problem; it does not calculate the Creatinine Clearance correctly for patients less than two years old". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2000-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens LANTIS 6.1 Oncology Information System with MEDOne Module, Linear Medical Accelerators; Product Numbers: 4503178, 5835991, 5851055; Product is manufactured by IMPAC Medical, Inc., Sunnyvale, CA. Medical charged-particle radiation therapy system.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- The product has a software problem; it does not calculate the Creatinine Clearance correctly for patients less than two years old.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- On 3/6/2009, a Safety Advisory Letter was distributed to all affected users, with a description of the problem and identifying the affected product. Customers were instructed not to use the LANTIS to calculate Creatine Clearance for any patient under two years of age. Customers should direct their questions with regards to the letter to their local Siemens Service organization. A Siemens Representative will update the systems.
- Quantity in commerce
- 17
- Distribution
- Worldwide Distribution
- Product codes and lots
- All units of these product numbers. Software version 6.1