Z-1394-2009

Roche cobas 6000 analyzer series (including cobas c 501 and cobas e 601 modules) Middleware Solutions...

Roche Diagnostics GmbH / initiated 2009-04-01 / Terminated / root cause: design

Roche Diagnostics Corp. began a recall of Roche cobas 6000 analyzer series (including cobas c 501 and cobas e 601 modules) Middleware Solutions Instrument Manager Server as follows: v.8.04.02.10... on . The reason FDA records is "Some data alarm flags do not transfer from the analyzer to the laboratory information system, since not all data flags are recognized by the instrument manager/middleware". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1394-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche cobas 6000 analyzer series (including cobas c 501 and cobas e 601 modules) Middleware Solutions Instrument Manager Server as follows: v.8.04.02.10, catalog no. 04785568001; v.8.04.03.01, catalog no. 04785568002; and v.8.06.02, catalog no. 05109019001. The device is used to interface hospital LIS systems with Roche c 6000 analyzers and allows order and result communication to occur between the two systems.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Some data alarm flags do not transfer from the analyzer to the laboratory information system, since not all data flags are recognized by the instrument manager/middleware.
Root cause tag
design
FDA root cause
Component design/selection
Action
Consignees were notified by "Urgent Medical Device Correction" letter dated April 1, 2009 and instructed to add error codes to the Instrument Manager driver and to contact the manufacturer of their LIS for instructions on adding data alarm codes to the LIS. Consignees were also instructed to complete the attached fax form after adding the error codes to their system and fax to 1-888-345-4690. Contact Roche Diagnostics Technical Support at 1-800-428-2336 for questions about information contained in the notification letter.
Quantity in commerce
498 of all products.
Distribution
Nationwide Distribution.
Product codes and lots
All units.