Z-1254-2009
AXIOM Sireskop SD Model numbers 8890415 and 8890407. The product is intended for use as radiographic and...
Siemens Medical Solutions USA, Inc. / initiated 2009-01-08 / Terminated / root cause: manufacturing process
Siemens Medical Solutions USA, Inc began a recall of AXIOM Sireskop SD Model numbers 8890415 and 8890407. The product is intended for use as radiographic and fluoroscopic imaging device on . The reason FDA records is "During interventional use and applications, the possibility exists for liquids to enter the systems". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1254-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- AXIOM Sireskop SD Model numbers 8890415 and 8890407. The product is intended for use as radiographic and fluoroscopic imaging device.
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- During interventional use and applications, the possibility exists for liquids to enter the systems.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- A Customer Safety Advisory Notice dated January 8, 2009 was issued to affected customers via Update Instructions AX069/08/S. This letter informed customers of the potential issues and provided instructions to avoid its occurrence. The firm has released an Addendum to the User Manual warning of the potential hazards if liquids penetrate the systems during interventional use and applications. Direct questions about this recall to Siemens Medical Solutions USA, Inc. by calling 1-610-219-6300.
- Quantity in commerce
- 240 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Serial numbers: 51010, 71136, 51007, 51008, 71019, 71080, 71172, 71067, 41012, 71015, 71038, 71102, 71104, 71169, 51006, 51011, 51024, 51028, 71026, 71039, 71083, 71117, 71119, 41027, 71120, 71121, 41028, 71137, 71138, 71149, 71158, 71161, 71167, 71168, 71170, 51060, 71160, 51026, 71030, 71059, 71033, 71037, 41006, 51052, 51064, 71001, 71082, 71114, 71141, 71148, 71152, 71156, 51046, 71181, 51013, 51044, 71075, 71098, 51054, 71020, 71021, 71074, 71088, 41026, 71135, 71140, 71055, 71006, 71017, 71042, 71073, 71162, 71174, 41003, 51049, 71053, 71128, 71131, 41022, 71096, 71099, 71129, 51009, 51036, 51045, 71069, 71087, 71144, 41008, 71014, 71016, 71028, 71109, 71130, 71004, 71051, 71090, 71157, 51029, 51030, 51069, 71009, 71011, 71034, 71035, 71126, 71127, 71145, 51003, 51016, 71022, 71029, 71164, 51043, 51070, 71061, 71064, 71110, 71116, 51001, 51057, 41014, 71124, 71133, 51012, 51050, 71070, 71142, 71095, 41011, 51047, 71023, 71076, 71081, 71084, 51041, 41010, 71032, 41029, 71002, 71065, 71105, 71115, 51005, 71068, 51017, 51018, 51023, 71043, 71046, 41020, 71091, 71092, 71101, 71108, 71139, 71146, 71150, 71155, 71178, 51015, 51059, 51066, 71013, 71057, 71062, 71097, 41013, 71024, 41023, 71086, 51037, 71010, 71106, 71112, 51065, 51068, 71008, 41017, 71078, 41024, 71079, 71111, 71018, 71147, 51002, 71040, 41002, 51019, 51020, 41005, 51033, 41007, 51040, 51048, 51051, 51055, 51061, 71003, 71007, 71012, 41018, 71025, 41019, 71045, 71063, 71100, 71103, 71113, 71122, 71125, 71153, 71154, 71159, 71163, 71171, 71173, 71177, 51004, 51021, 51025, 51058, 71048. 71058, 71060, 71072, 41001, 51022, 51031, 51035, 51039, 71027, 71175, 51063, 71005, 71066, 71089, 71165, 71166, 71123, 71151, and 51056.