Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit NL950-V and Integra NeuroSciences...
Integra Neuro Sciences began a recall of Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit NL950-V and Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring... on . The reason FDA records is "Due to a dimensional error, the diameter of the stylet provided with the device may be too large to insert into the lumen of the catheter tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2053-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Natus Medical Incorporated
- Product
- Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit NL950-V and Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit with Ventricular Drainage System NL950-VC. This intracranial pressure (ICP) catheter is indicated for use on patients that require continuous invasive intracranial pressure monitoring and/or cerebrospinal drainage. Access is achieved through a tunneled technique using the included trocar.
- Product code
- JXG Shunt, Central Nervous System And Components
- Reason for recall
- Due to a dimensional error, the diameter of the stylet provided with the device may be too large to insert into the lumen of the catheter tip.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- An "Urgent-Product Recall Notification" letter dated March 23, 2009 was issued to consignees. The letter described the issue and instructed customers to examine their inventory immediately to determine whether they have any NL950V - Ventrix Ventricular Tunneling Pressure Monitoring Kits, or NL950VC - Ventrix Ventricular Tunneling Pressure Monitoring Kits with Ventricular Drainage System. Customers were further instructed not to use the above recalled product. Call Integra Customer Service at 1-800-654-2873, and then select 5 for 'Returns and Repairs' to obtain a Returned Material Authorization number, and return the product as directed. Direct questions about the recall to Integra NeuroSciences by calling 1-858-455-1115, extension185.
- Quantity in commerce
- 274 kits
- Distribution
- Worldwide Distribution -- US (states of AR, CA, OH, TX, CT, NY, LA, FL, NY, IL, and VA), Belgium and Malaysia.
- Product codes and lots
- NL950-V - Lot Numbers: 30500R130474 (exp. 8/31/2011); 30500R134098 (exp. 10/31/2011); 30500R139222 (exp. 11/30/2011); 30500R143588 (exp. 1/31/2012); 30500R146383 (exp. 1/31/2012); 30500R147342 (exp. 2/29/2012); and 30500R147899 (exp. 2/29/2012) and NL950-VC - Lot Numbers: 138999 (exp. 9/30/2011); and 142754 (exp. 12/31/2011).