Z-1422-2009

EP-4 Computerized Cardiac Stimulator; SJM Part Numbers: 09-1527-0000 / 09-1527-0002; EP MedSystems, Inc. 575...

Abbott Medical / initiated 2009-03-23 / Terminated / root cause: design

St. Jude Medical began a recall of EP-4 Computerized Cardiac Stimulator; SJM Part Numbers: 09-1527-0000 / 09-1527-0002; EP MedSystems, Inc. 575 Route 73 North, Units D, West Berlin, NJ 08091... on . The reason FDA records is "Non-compliance with IEC Standard 60601-1 Clause 57.6 relative to fusing of AC input line. The neutral line is not fused. The hot line in fused. Clause 57.6 calls for both lines to be fused". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1422-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
EP-4 Computerized Cardiac Stimulator; SJM Part Numbers: 09-1527-0000 / 09-1527-0002; EP MedSystems, Inc. 575 Route 73 North, Units D, West Berlin, NJ 08091. Indicated for use during clinical electrophysiology procedures.
Product code
JOQ Generator, Pulse, Pacemaker, External Programmable (For Electrophysiological Studies Only)
Reason for recall
Non-compliance with IEC Standard 60601-1 Clause 57.6 relative to fusing of AC input line. The neutral line is not fused. The hot line in fused. Clause 57.6 calls for both lines to be fused.
Root cause tag
design
FDA root cause
Device Design
Action
All foreign customers were sent Urgent Product Safety Information letters on March 23, 2009. Letters described units affected, the reason for action, and the actions to take to eliminate risk. It is recommended that the information in the letter be circulated to all users, risk managers and the biomedical engineering department responsible for maintaining and inspecting electrical equipment located within their institution. The letter also stated that a SJM representative will be in touch to perform the EP-4 Stimulator fuse modification. Customers are to complete and return the attachment.
Quantity in commerce
694 total (US, Canada, and OUS)
Distribution
Worldwide distribution.
Product codes and lots
Systems Part Numbers: EP-4I-2-110; EP-4I-2-220; EP-4I-4-110; EP-4-220: WM-01M-241; WM-01M-560; WM-01M-1200; WM-02M-241; WM-02M-560; and WM-02M-1200.