Z-1728-2009

Hologic Delphi series Bone Densitometer Software versions 12.6, 2.3., or 2.4 Indicated for the estimation of...

Hologic, Inc. / initiated 2009-04-07 / Terminated / root cause: design

Hologic, Inc. began a recall of Hologic Delphi series Bone Densitometer Software versions 12.6, 2.3., or 2.4 Indicated for the estimation of bone mineral density (BMD), comparison of... on . The reason FDA records is "System may not turn off at the end of the x-ray and cause additional x-ray exposure". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1728-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hologic, Inc.
Product
Hologic Delphi series Bone Densitometer Software versions 12.6, 2.3., or 2.4 Indicated for the estimation of bone mineral density (BMD), comparison of measurable variables obtained from a given QDR scan to a database of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics.
Product code
KGI Densitometer, Bone
Reason for recall
System may not turn off at the end of the x-ray and cause additional x-ray exposure
Root cause tag
design
FDA root cause
Component design/selection
Action
Hologic issued an Urgent: Medical Device Recall customer notification letter, dated April 17, 2009, by certified mail to inform customers of the problem. With the notification letter, customers will receive the following: -Software version Apex 12.6.2, 2.3.2, or 2.4.2 -Instructions to upgrade their systems with the provided software and to verify the new version on the system as a confirmation that the software was successfully loaded. -New Operator's Manual for installations where the operating characteristics of the system have changed due to the software upgrade. For Customers with systems with software versions other than 12.6, 2.3, or 2.4: Customers were asked to upgrade their QDR software, return their old QDR software disk, and complete and return to Hologic the enclosed Confirmation of Notification.
Quantity in commerce
1914 units total
Distribution
Worldwide Distribution -- USA, Canada, Europe, Austria, Belgium, Bulgaria, Czech Republic, Denmark, France, Germany, Great Britain, Greece, Hungary, Iceland, Italy, Latvia, Lithuania, Macedonia, Netherlands, Norway, Poland, Portugal, Romania, Russia, Serbia, Slovak Republic, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine, Uzbekistan, Algeria, Arab Emirates, Argentina, Australia, Bahrain, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Egypt, Hong Kong, Indonesia, India, Iran, Iraq, Israel, Japan, Jordan, Kenya, Korea, Kuwait, Lebanon, Malaysia, Morocco, Mexico, Pakistan, Peru, Puerto Rico, South Africa, Saudia Arabia, Singapore, Syria, Taiwan, Thailand, Tunisia, Venezuela, and Vietnam.
Product codes and lots
S/N: 70638A 70756 70387C 71202W, and  70089W.