Post Medical Valves VO870A-Y and VO870-A-Y-T PSI 2015, PSI 2015, CGA 870, Inlet .750 STDS, Gas services...
Intermed Gas Products Corporation began a recall of Post Medical Valves VO870A-Y and VO870-A-Y-T PSI 2015, PSI 2015, CGA 870, Inlet .750 STDS, Gas services (Oxygen), PSI 2015, Aluminum Cylinders, InterMed Gas... on . The reason FDA records is "Firm received reports from customers that valves malfunctioned when installation and/or filling process. Customers forwarded pictures of the defect where the top portion of the packing nut broke off from the threaded...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2011-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intermed Gas Products Corporation
- Product
- Post Medical Valves VO870A-Y and VO870-A-Y-T PSI 2015, PSI 2015, CGA 870, Inlet .750 STDS, Gas services (Oxygen), PSI 2015, Aluminum Cylinders, InterMed Gas Products LLC Industrial Division, 2056 Vista Parkway West Palm Beach, FL 33411.
- Product code
- ECX Cylinder, Compressed Gas, And Valve
- Reason for recall
- Firm received reports from customers that valves malfunctioned when installation and/or filling process. Customers forwarded pictures of the defect where the top portion of the packing nut broke off from the threaded portion of the net; which may cause the valve to fail and result in uncontrolled release of oxygen when the valve is in the open position.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Intermed Gas Products Corporation notified customers of the affected product by phone beginning March 27, 2009 and followed up with a formal "Product Recall Notice" letter dated March 31, 2009. All affected product should be located and returned to the firm for replacement. For further information, contact Intermed Gas Products at 1-561-483-1364.
- Quantity in commerce
- 8,767
- Distribution
- Nationwide Distribution -- FL, SC, TX, MO, TN, NC, NY, GA, MN, AZ.
- Product codes and lots
- Batch Numbers: 08-05-3B and 09-01-1B.