ILLICO Posterior Thoracolumbar Access Retractor, Part Number: 73520 The ILLICO Posterior Thoracolumbar Access...
Alphatec Spine, Inc. began a recall of ILLICO Posterior Thoracolumbar Access Retractor, Part Number: 73520 The ILLICO Posterior Thoracolumbar Access Retractor is an instrument that is intended for... on . The reason FDA records is "The recall was initiated after Alphatec Spine received a complaint on the ILLICO Posterior Thoracolumbar access retractor. It was reported that one of the retractor arms in the retractor broke during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0498-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alphatec Spine, Inc.
- Product
- ILLICO Posterior Thoracolumbar Access Retractor, Part Number: 73520 The ILLICO Posterior Thoracolumbar Access Retractor is an instrument that is intended for use in a minimally invasive approach for surgical procedures. The retractor allows retraction of tissue through combination of articulating blades. Device will serve to provide access for surgeons to surgical site.
- Product code
- GAD Retractor
- Reason for recall
- The recall was initiated after Alphatec Spine received a complaint on the ILLICO Posterior Thoracolumbar access retractor. It was reported that one of the retractor arms in the retractor broke during surgery.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On January 28, 2009 Alphatec Spine contacted their domestic consignees via telephone and the international consignee was contacted via telephone and electronic mail. Alphatec Spine's telephone call consisted of informing the consignees of the product affected, the problem, what actions needed to be taken and what action Alphatec Spine would take to arrange for pick-up of the affected products.
- Quantity in commerce
- 14
- Distribution
- AZ, CA, OR, TX, and Tokyo, Japan.
- Product codes and lots
- Lot number: 5271901