Z-1324-2009
Square Nut MR Safe, Catalog Number 03.311.060. The Square Nut MR Safe is used in conjunction with Synthes...
Synthes GmbH / initiated 2009-02-09 / Terminated / root cause: manufacturing process
Synthes Spine began a recall of Square Nut MR Safe, Catalog Number 03.311.060. The Square Nut MR Safe is used in conjunction with Synthes Distraction Osteogenesis Ring Systems. It can be used... on . The reason FDA records is "Numbers are etched incorrectly on the product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1324-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Square Nut MR Safe, Catalog Number 03.311.060. The Square Nut MR Safe is used in conjunction with Synthes Distraction Osteogenesis Ring Systems. It can be used to secure components together or as a means of bone distraction/compression. Each side of the Square Nut is etched with a number, 1-2-3-4, which can be referenced during distraction/compression. Distraction of bone typically occurs a a rate of 1mm/day. The patient may be instructed to turn the nuts 1/4 turn four times per day, accomplishing 0.25mm per 1/4 turn. The etched numbers offer reference points to the surgeon who instructs the patient on distraction technique.
- Product code
- KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
- Reason for recall
- Numbers are etched incorrectly on the product.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- An "Urgent: Medical Device Recall" letter dated February 6, 2009 was sent to all affected Synthes Trauma Sales Consultants via USPS certified mail on February 09, 2009 and affected In House employees were notified via return receipt email on February 09, 2009. All affected hospitals were notified of the recall through a notification letter mailed via USPS certified mail on February 09, 2009. The notification letter described the affected product, issue and actions for recipients. Direct questions about the recall to Synthes Spine by calling 800-620-7025, Extension 5455 or 610-719-5455.
- Quantity in commerce
- 1,137 units
- Distribution
- Worldwide Distribution -- US (states of AR, CA, CO, CT, FL, IA, IL, LA, MD, ME, MI, MN, MS, NJ, NY, OK, PA, SC, TX, and UT) and country of Switzerland.
- Product codes and lots
- Lot Numbers: UR76230, US89403, US94671, US97826, 5674035, US90907, US94396 AND US97022.