Z-1504-2009
AccuSure, U-100 31 Gauge, 1cc Insulin Syringe, 5/16" (8mm) Needles, Short Needle, packaged 10 syringes per...
Vintage Pharmaceuticals LLC dba Qualitest Pharmaceuticals / initiated 2009-04-09 / Terminated / root cause: manufacturing process
Vintage Pharmaceuticals LLC DBA Qualitest Pharmaceuticals began a recall of AccuSure, U-100 31 Gauge, 1cc Insulin Syringe, 5/16" (8mm) Needles, Short Needle, packaged 10 syringes per package, 10 packages per carton, NDC 0603-7002-2 on . The reason FDA records is "Needles separate from the barrel of the syringes". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1504-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Vintage Pharmaceuticals LLC dba Qualitest Pharmaceuticals
- Product
- AccuSure, U-100 31 Gauge, 1cc Insulin Syringe, 5/16" (8mm) Needles, Short Needle, packaged 10 syringes per package, 10 packages per carton, NDC 0603-7002-2
- Product code
- FMF Syringe, Piston
- Reason for recall
- Needles separate from the barrel of the syringes.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 04/09/2009, Qualitest issued a press release and sent out recall notifications to their consignees by US Mail. The recall was extended to the User level. Qualitest will send a second letter to all non-responders after thirty days. Thirty days after the second letter, Qualitest will perform a recall effectiveness check by telephone on a percentage of non-responders (in accordance with FDA request).
- Quantity in commerce
- 559,500 syringes
- Distribution
- Nationwide.
- Product codes and lots
- Lot A07001, exp 07/2012
FIRST ARTICLE / WEEKLY