Z-1155-2009
GE Healthcare Fluorostar 7900 Mobile Fluoroscopy System
GE Healthcare / initiated 2009-02-19 / Terminated
Ge Healthcare began a recall of GE Healthcare Fluorostar 7900 Mobile Fluoroscopy System on . The reason FDA records is "The alarm timer is reset after each exposure and therefore never reaches the five minute alarm state unless one exposure exceeds 5 minutes in duration". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1155-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Fluorostar 7900 Mobile Fluoroscopy System
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- The alarm timer is reset after each exposure and therefore never reaches the five minute alarm state unless one exposure exceeds 5 minutes in duration.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- GE Healthcare sent a customer notification out to all sites that have potentially affected systems. GEHC Surgery will also install a software correction into the affected systems in order to resolve this issue at no charge to the customer. Until the GEHC OEC Fluorostar 7900 has received the corrective software installed, they are encouraging facilities to exercise caution in monitoring the amount of cumulative fluoroscopy time when the affected system is in use.
- Quantity in commerce
- 554 units
- Distribution
- Worldwide Distribution
- Product codes and lots
- n/a