Z-1428-2009
Ultrasite IV Set for Outlook Safety Infusion System with Universal Spike, Pressure Limited Check Valve, 3...
B Braun Medical Inc / initiated 2009-03-27 / Terminated
B. Braun Medical, Inc. began a recall of Ultrasite IV Set for Outlook Safety Infusion System with Universal Spike, Pressure Limited Check Valve, 3 Ultrasite Injection Sites, Backcheck Valve, and... on . The reason FDA records is "Device defect is occlusion/no flow". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1428-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- B Braun Medical Inc
- Product
- Ultrasite IV Set for Outlook Safety Infusion System with Universal Spike, Pressure Limited Check Valve, 3 Ultrasite Injection Sites, Backcheck Valve, and Spin-Lock Connector, Catalog # 352520. For use with the Horizon NXT Pump, and OutLook Safety Infusion System.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Device defect is occlusion/no flow.
- FDA root cause
- Mixed-up of materials/components
- Action
- An Urgent - Medical Device Recall notice was issued via US Postal Service Certified Mail with registered return receipt mail or FedEx to all customer/distributors in receipt of affected product on 3/27/09. Direct customers received letters addressed to Materials Management that informed them of this recall and provided instructions for disposition of affected product in their inventory. Distributors received the direct customer recall letter along with a cover letter outlining their responsibilities to execute the recall and notify their customers as requested. Distributors are to immediately discontinue further distribution of the affected product, determine their customers' inventory, and have their customers return the product to them. The attached "Product Removal Acknowledgement" should be used to indicate the specifics of the combined distributor and customer inventories. The Customer Support Department should be contacted to arrange for return. A Customer Support Representative will provide instructions for handling the affected product and assist with replacement product.
- Quantity in commerce
- 13,008 units
- Distribution
- Worldwide Distribution -- USA and Canada.
- Product codes and lots
- Lot #'s 61040841 (exp 2/28/2012) and 61040842 (exp 1/31/2012).