Z-1502-2009
Arrow OTW Percutaneous Thrombolytic Device (PTD) for fistulae and dialysis grafts
Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2009-02-11 / Terminated / root cause: design
Arrow International Inc began a recall of Arrow OTW Percutaneous Thrombolytic Device (PTD) for fistulae and dialysis grafts on . The reason FDA records is "Leakage: Material not lipid resistant, may crack and/or leak. Use of this device with lipid containing solutions or 70% isopropyl alcohol may cause leakage or air embolism". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1502-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Arrow OTW Percutaneous Thrombolytic Device (PTD) for fistulae and dialysis grafts
- Product code
- DXE Catheter, Embolectomy
- Reason for recall
- Leakage: Material not lipid resistant, may crack and/or leak. Use of this device with lipid containing solutions or 70% isopropyl alcohol may cause leakage or air embolism.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Arrow International issued a letter dated 2/18/09 to all customers informing them of the problem and if a lipid resistant stopcock is required to contact the firm for replacements. New labels state: Warning - do not use with lipid containing solutions or 70% isopropyl alcohol, as these solutions may cause leakage or air embolism.
- Quantity in commerce
- 900,000 total, all products
- Distribution
- Worldwide distribution: USA, Argentina, Australia, Bahamas, Belarus, Belgium, Brazil, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Latvia, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Singapore, Slovakia, South Africa, South Korea, Spain, Sri lanka, Sweden, Taiwan, Thailand, Tunisia, Turkey, UK, Ukraine, and Venezuela.
- Product codes and lots
- All lots of Product codes listed in Z-1489-2009.