Welch Allyn AED 10 automatic external defibrillator and MRL Jumpstart (collectively "AED 10")
Welch Allyn Protocol, Inc began a recall of Welch Allyn AED 10 automatic external defibrillator and MRL Jumpstart (collectively "AED 10") on . The reason FDA records is "Possible for users to misunderstand signals. The original product labeling told users that a flashing low battery indicator signals that the battery is low and requires changing soon. If a user experiences a flashing...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1656-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Welch Allyn Protocol, Inc.
- Product
- Welch Allyn AED 10 automatic external defibrillator and MRL Jumpstart (collectively "AED 10")
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Possible for users to misunderstand signals. The original product labeling told users that a flashing low battery indicator signals that the battery is low and requires changing soon. If a user experiences a flashing low battery status indicator and does not understand that the device can continue to be used, the user might choose not to continue operating the device, thereby causing therapy to
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- The firm has developed expanded labeling for users and will revise the user manuals for software versins 2.03, 2.04, 2.05, and 2.06. Customers receiving the updated manuals will be sent a cover letter directing them to discard the original user manual and replace it with the updated manual. The April 24, 2009, cover letter will highlight the clarification of directions for use regarding what to do when a low battery indicator is displayed.
- Quantity in commerce
- 46740 units
- Distribution
- Worldwide distribution: USA, Guam, Puerto Rico, Canada, Mexico, Andorra, UAE, Afghanistan, Australia, Azerbaijan, Belgium, Belarus, Bermuda, Bahrain, Brazil, British West Indies, Switzerland, China, Cyprus, Czech Republic, Germany, Denmark, Algeria, Egypt, Spain, France, French Guiana, French Lesser Antilles, Greece, Haiti, Hong Kong, Hungary, Ireland, Israel, India, Iraq, Italy, Japan, Kenya, Kyrgyz, Lebanon, Lithuania, Luxembourg, Latvia, Monaco, Malaysia, Netherlands, New Zealand, Phillipines, Pakistan, Poland, Portugal, Reunion Island, Russia, Saudi Arabia, Singapore, South Korea, Taiwan, United Kingdom, Uraguay, Yemen, and South America.
- Product codes and lots
- software versions 2.03, 2.04, 2.05, and 2.06