Z-1391-2009

Triage Cardiac Panel, Biosite Catalog Number 97000HS, 25 individually pouched devices in a labeled kit box...

Alere San Diego, Inc. / initiated 2009-04-27 / Terminated / root cause: manufacturing process

Biosite Inc began a recall of Triage Cardiac Panel, Biosite Catalog Number 97000HS, 25 individually pouched devices in a labeled kit box. Fluorescence immunoassay kit to be used with the... on . The reason FDA records is "The recall was initiated because the use of the affected lot may lead to reporting falsely low values for troponin I and myoglobin, possibly resulting in missed or incorrect diagnosis". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1391-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alere San Diego, Inc.
Product
Triage Cardiac Panel, Biosite Catalog Number 97000HS, 25 individually pouched devices in a labeled kit box. Fluorescence immunoassay kit to be used with the Triage Meter for the quantitative determination of myoglobin, creatine kinase MB fraction (CK-MB), and troponin I in EDTA whole blood and plasma specimens. These three cardiac biomarkers are used as an aid in the diagnosis of myocardial infarction (injury).
Product code
MMI Immunoassay Method, Troponin Subunit
Reason for recall
The recall was initiated because the use of the affected lot may lead to reporting falsely low values for troponin I and myoglobin, possibly resulting in missed or incorrect diagnosis.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The recall was initiated with "Urgent Medical Device Recall" notices, dated 4/28/2009, with attached "Customer/Distributor Verification Form" being sent to the affected customers via fax, e-mail or direct mail on 04/28/2009. Customers were informed of the reason for the recall and instructed to immediately discontinue all use of the listed products and discard all affected product in accordance with your their regulations. The customers were give two options; "Option 1: Complete and FAX the enclosed Verification Form within 10 days to confirm your receipt of this notice, to indicate the number of kits discarded from your inventory and product replacement will be sent to you. Option 2: Contact inverness Medical Technical Service at 1-877-441-7440 and select option 2 from the phone prompt. A representative will confirm the number of kits discarded from your inventory and arrange for replacement product to be sent to you." Customers were also provided telephone, fax and e-mail contact information for Biosite should they have any questions about the information contained in the recall notification.
Quantity in commerce
401 kits
Distribution
Nationwide Distribution
Product codes and lots
Lot Number W44467B