TransFx External Fixation System Drill Bit, Quick Connect, 3.0 mm diameter, 180 mm length, nonsterile...
Zimmer Inc. began a recall of TransFx External Fixation System Drill Bit, Quick Connect, 3.0 mm diameter, 180 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-58. TransFx drill bits can... on . The reason FDA records is "Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1534-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- TransFx External Fixation System Drill Bit, Quick Connect, 3.0 mm diameter, 180 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-58. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.
- Product code
- LXT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite
- Reason for recall
- Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Distributors and consignees were notified by an Urgent: Device Recall letter dated 5/1/09. Consignees were advised to stop using the device and quarantine it immediately. Distributors were instructed to carry out a physical count of all affected product and record the data on the Inventory Return Certification Form; fax a copy of the completed form to Zimmer, Inc.; and return recalled product along with the Inventory Return Certification Form to the firm.
- Quantity in commerce
- 707
- Distribution
- Worldwide Distribution -- USA, Australia, Canada, El Salvador, England, Germany, Latin America, and Singapore.
- Product codes and lots
- Lots 53007600, 53024800, 56454669, 70185700, 70221300, 70227300, 70249300, 70249400, 70250100 and 70254900.