Z-1424-2009

Medtronic, Transfer Spike, Blood Processing Accessory, REF BT945, Sequestration Kits

Medtronic Perfusion Systems / initiated 2009-02-25 / Terminated / root cause: design

Medtronic Cardiovascular Revascularization & Surgical Therap began a recall of Medtronic, Transfer Spike, Blood Processing Accessory, REF BT945, Sequestration Kits on . The reason FDA records is "Sterility may be compromised. Non-vented caps that connect to the 3-way stopcock attached to the transfer spike could prevent Ethylene Oxide from contacting the inside surfaces of the stopcock during sterilization...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1424-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Perfusion Systems
Product
Medtronic, Transfer Spike, Blood Processing Accessory, REF BT945, Sequestration Kits
Product code
DWE Tubing, Pump, Cardiopulmonary Bypass
Reason for recall
Sterility may be compromised. Non-vented caps that connect to the 3-way stopcock attached to the transfer spike could prevent Ethylene Oxide from contacting the inside surfaces of the stopcock during sterilization, leading to a potentially non-sterile product. Use of the product may cause serious health consequences from potential sepsis/infection and the associated remote risk of death.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were sent a Medtronic "Urgent Medical Device Recall Notice" letter dated March 5, 2009. The letter was addressed to Risk Manager. The letter described the product and problem and requested the consignee to complete, sign and return the certificate acknowledgement.
Quantity in commerce
20 units
Distribution
Distribution to one medical facility in MN.
Product codes and lots
Lot 10279571, Use By 2012-01-31