HACH SteriChek Residual Peroxide Reagent Strips, 5/100-strip bottles per kit, Kit Part #811905, Bottle part...
Hach Company / initiated 2009-04-21 / Terminated / root cause: labeling and UDI
Hach Co began a recall of HACH SteriChek Residual Peroxide Reagent Strips, 5/100-strip bottles per kit, Kit Part #811905, Bottle part #1905LB, Hach Co., Ames, IA. SteriChek Residual... on . The reason FDA records is "Some of the reagent strip bottles contain the incorrect strips". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2016-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hach Company
- Product
- HACH SteriChek Residual Peroxide Reagent Strips, 5/100-strip bottles per kit, Kit Part #811905, Bottle part #1905LB, Hach Co., Ames, IA. SteriChek Residual Peroxide Reagent Strips provide a convenient means of measuring the concentration of peroxide remaining in water being used to rinse out dialysate lines and dialyzers following disinfection with peroxide/peracetic acid.
- Product code
- LIF Dialyzer Reprocessing System
- Reason for recall
- Some of the reagent strip bottles contain the incorrect strips
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Hach Company issued an "Urgent Medical Device Recall" letter dated April 21, 2009 explaining the reason for recall and requested product remaining in inventory be returned. Subrecall to the end-user level was requested. A Return Form was also enclosed to be completed to indicate the amount of product being returned. Between April 22-23, 2009, the firm telephoned their customers to assure they received the recall letter. For further information, contact Hach Customer Service at 1-800-548-4381 extension 3560.
- Quantity in commerce
- 275 kits
- Distribution
- Nationwide Distribution -- NJ, CA, IN, PA, NV, CA, and ME.
- Product codes and lots
- Part Number: 811905, Lot Number: A8350; and Part Number: 1905LB, Lot Numbers: 8350 and 8345.