Z-1930-2009
Dimension Flex Reagent Cartridge C-Reactive Protein Extended Range (RCRP). In vitro diagnostic test intended...
Siemens Healthcare Diagnostics Inc. / initiated 2008-04-15 / Terminated / root cause: labeling and UDI
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Flex Reagent Cartridge C-Reactive Protein Extended Range (RCRP). In vitro diagnostic test intended for the quantitative determination of CRP in human... on . The reason FDA records is "falsely elevated results in EDTA plasma samples". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1930-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Flex Reagent Cartridge C-Reactive Protein Extended Range (RCRP). In vitro diagnostic test intended for the quantitative determination of CRP in human serum and plasma (lithium heparin). Measurement of C-Reactive Protein is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.
- Product code
- DCN System, Test, C-Reactive Protein
- Reason for recall
- falsely elevated results in EDTA plasma samples
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice on April 15, 2008 informing their customers that the EDTA plasma specimen claim had been rescinded. Current lots in the firm's inventory have been put on hold and reworked with an Alert Card rescinding the EDTA plasma claim. Subsequent lots use a revised instruction for use stating "Do not use EDTA plasma" in the Specimen Collection and Handling Section. For further questions, contact Siemens Healthcare Diagnostics at 1-302-631-6564.
- Quantity in commerce
- 25,760 cartons
- Distribution
- Worldwide Distribution -- US, Austria, Bosnia, Czech Republic, Hungary, Slovakia, Serbia, Denmark, Estonia, Latvia, Lithuania, Norway, Sweden, Belgium, Cyprus, Canary Islands, France, Great Britain, Germany, Ireland, Italy, Netherlands, Poland, Russia, Portugal, Saudi Arabia, Spain, South Africa, Switzerland, Turkey, Yemen, Bahamas, Brazil, Canada, Mexico, Australia, Bangladesh, China, Hong Kong, Japan, Malaysia, Pakistan, and Taiwan.
- Product codes and lots
- Catalog Number DF34; Lot Numbers: CE8324 exp 11/19/2008, BA8349 exp 12/14/2008, FB9003 exp1/3/2009, BC9017 exp1/17/2009, EC9021 exp1/21/2009, DB9032 exp2/1/2009, and DC9045 exp2/14/2009.