Zimmer NexGen Complete Knee Solution Legacy Knee - Posterior Stabilized LPS-Flex Femoral Component Precoat...
Zimmer Inc. began a recall of Zimmer NexGen Complete Knee Solution Legacy Knee - Posterior Stabilized LPS-Flex Femoral Component Precoat, Size D left , sterile, Zimmer, Warsaw, IN; REF... on . The reason FDA records is "The implant surface may not have been polished adequately, which could result in increased wear and polyethylene debris generation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1520-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer NexGen Complete Knee Solution Legacy Knee - Posterior Stabilized LPS-Flex Femoral Component Precoat, Size D left , sterile, Zimmer, Warsaw, IN; REF 00-5960-014-51.
- Product code
- NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing
- Reason for recall
- The implant surface may not have been polished adequately, which could result in increased wear and polyethylene debris generation.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Consignees were notified by phone beginning on 4/17/09 and email and letter notifications were sent on 4/22/09. A dear surgeon letter entitled Urgent: Medical Device Recall, dated April 28, 2009, is being sent to implanting surgeons recommending patient monitoring. Consignees are to return the response form indicating their receipt of the notification. Questions or concerns should be directed to (800) 846-4637 or www.zimmer.com.
- Quantity in commerce
- 5
- Distribution
- Worldwide Distribution -- USA including states of Arizona, California, Colorado, Hawaii, Illinois, Indiana, Michigan, Minnesota, Mississippi, New Jersey, New Mexico, New York, Ohio, Oregon, and Pennsylvania and country of Canada.
- Product codes and lots
- Lot 61208295.