Z-1477-2009

Zimmer Gender Solutions Natural Knee Flex System, Gender Solutions Male, Femoral Component, nonporous, size...

Zimmer GmbH / initiated 2009-04-29 / Terminated / root cause: manufacturing process

Zimmer Inc. began a recall of Zimmer Gender Solutions Natural Knee Flex System, Gender Solutions Male, Femoral Component, nonporous, size 4, left, sterile, Zimmer, Warsaw, IN; Product... on . The reason FDA records is "The tyvek lid, which is intended to assure sterility until the device is transported to the operating room, may be missing from the inner tray". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1477-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Gender Solutions Natural Knee Flex System, Gender Solutions Male, Femoral Component, nonporous, size 4, left, sterile, Zimmer, Warsaw, IN; Product Number: 00-5410-017-01. Knee implant used to replace the distal surface of the human femur in total knee arthroplasty surgery. For use in Non-inflammatory Degenerative Joint Disease or Inflammatory Joint Disease, deformity, moderate contracture or failed previous surgery. Product is indicated for uncemented or cemented use.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The tyvek lid, which is intended to assure sterility until the device is transported to the operating room, may be missing from the inner tray.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Zimmer, Inc. issued an "Urgent: Device Recall" letter dated April 29, 2009 informing all consignees of the affected product. Users were instructed to locate and quarantine all unused devices and ship them to the firm with a completed Inventory Return Certification form. Return Certification forms are also to be sent via fax to Zimmer, Inc. at 1-574-372-4265. For further questions, contact Zimmer, Inc. at 1-800-613-6131.
Quantity in commerce
14
Distribution
Worldwide Distribution -- US including states of FL, MN, MS, NY, OH, TN, TX and WI and Switzerland.
Product codes and lots
Lot Number: 60998487.