Siemens Stratus CS Acute Care Troponin I CalPak. In vitro product intended to be used for calibration of the...
Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Stratus CS Acute Care Troponin I CalPak. In vitro product intended to be used for calibration of the cardiac troponin I method on the Stratus CS... on . The reason FDA records is "The firm received and confirmed that some lots of Cardiac Troponin I (cTnl) CalPaks (CCTNI-CR) may exhibit calibration issues that cause lower than typical signal during calibration with the current population of cTnl...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1555-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens Stratus CS Acute Care Troponin I CalPak. In vitro product intended to be used for calibration of the cardiac troponin I method on the Stratus CS analyzer.
- Product code
- MMI Immunoassay Method, Troponin Subunit
- Reason for recall
- The firm received and confirmed that some lots of Cardiac Troponin I (cTnl) CalPaks (CCTNI-CR) may exhibit calibration issues that cause lower than typical signal during calibration with the current population of cTnl TestPaks (CCTNI). This issue may result in failed calibrations due to "Slope Error" messages or falsely elevated patient sample and quality control (QC) results on calibrations that
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- Siemens Healthcare Diagnostics, Inc. issued an Urgent Field Safety Notice dated April 29, 2008 to all affected customers instructing them to discontinue use of 3 CalPak lots and reduce the shelf-life for the remaining lots. For further information, contact Siemens Healthcare Diagnostics, Inc. at 1-302-631-7672.
- Quantity in commerce
- 4415 cartons
- Distribution
- Worldwide Distribution - including United States, United Arab Emirates, Austria, Bermuda, Belize, Canada, Germany, Egypt, India, Japan, Saudi Arabia and Bengal.
- Product codes and lots
- Lot Numbers: 637218002 (expired immediately), 637281002 (expired immediately), 637350002 (expired immediately), 638035002 (expired June 4, 2008); and 638105002 (expired August 14, 2008).