Z-1416-2009

Cook Medical Tao Brush I.U.M.C. Endometrial Cytology/Histology Sampler (Tao Brush), 9.0Fr/26 cm, 3.5 cm...

Cook Ob/Gyn / initiated 2009-04-02 / Terminated / root cause: manufacturing process

Cook OB/Gyn, Inc. began a recall of Cook Medical Tao Brush I.U.M.C. Endometrial Cytology/Histology Sampler (Tao Brush), 9.0Fr/26 cm, 3.5 cm, sterile, Cook Ob/Gyn, Spencer, IN; REF J-ES-090500-BL... on . The reason FDA records is "There is an ink on the brush which has not been tested for use with the Tao Brush". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1416-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Ob/Gyn
Product
Cook Medical Tao Brush I.U.M.C. Endometrial Cytology/Histology Sampler (Tao Brush), 9.0Fr/26 cm, 3.5 cm, sterile, Cook Ob/Gyn, Spencer, IN; REF J-ES-090500-BL. Used to obtain endometrial samples for both cytology and histology. Intended for one-time use.
Product code
HFE Brush, Endometrial
Reason for recall
There is an ink on the brush which has not been tested for use with the Tao Brush.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The consignee was notified by an Urgent: Recall Notice letter dated 4/2/09. The customer is to forward the notice to all personnel that need to be aware of it; pull from their stock and isolate any affected product; and complete the form. Upon receipt of the form, Cook Women's Health will issue a Return Goods Authorization number and the customer should return the affected products via FedEx using the Return Goods Authorization number.
Quantity in commerce
15,396
Distribution
Nationwide Distribution.
Product codes and lots
Lots UF1793872 through UF1793877, UF1811910 through UF1811913, UF1811916, UF1811917, UF1811917S, UF1811918, UF1811919, UF1816687 and UF1816688.