Z-1479-2009

Philips Avalon Fetal Monitor FM30; Model Number: M2703A

Philips North America LLC / initiated 2009-05-12 / Terminated / root cause: software

Philips Healthcare Inc. began a recall of Philips Avalon Fetal Monitor FM30; Model Number: M2703A on . The reason FDA records is "Interruption of power-up/boot sequence may cause the paper scale and recorder speed settings of the monitor to revert to factory default may lead to misinterpretation of the fetal trace by the user and to potential...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1479-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Avalon Fetal Monitor FM30; Model Number: M2703A.
Product code
HGM System, Monitoring, Perinatal
Reason for recall
Interruption of power-up/boot sequence may cause the paper scale and recorder speed settings of the monitor to revert to factory default may lead to misinterpretation of the fetal trace by the user and to potential incorrect clinical decisions
Root cause tag
software
FDA root cause
Software change control
Action
Philips Healthcare issued an "Urgent-Medical Device Correction" notification dated May 15, 2009 to Customers in the United States via UPS and notifications outside the US will be managed by Philips representatives in each affected geography. Customers were asked to follow the guidelines in the Action to be Taken by Customer/User section of the Correction Notice until they receive their software upgrade. Questions can be made to Philips response center at 1-800-722-9377.
Quantity in commerce
4,788 units
Distribution
Worldwide Distribution
Product codes and lots
Units with serial numbers in the range of DE53001002 through DE53107383.