Z-1479-2009
Philips Avalon Fetal Monitor FM30; Model Number: M2703A
Philips North America LLC / initiated 2009-05-12 / Terminated / root cause: software
Philips Healthcare Inc. began a recall of Philips Avalon Fetal Monitor FM30; Model Number: M2703A on . The reason FDA records is "Interruption of power-up/boot sequence may cause the paper scale and recorder speed settings of the monitor to revert to factory default may lead to misinterpretation of the fetal trace by the user and to potential...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1479-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Avalon Fetal Monitor FM30; Model Number: M2703A.
- Product code
- HGM System, Monitoring, Perinatal
- Reason for recall
- Interruption of power-up/boot sequence may cause the paper scale and recorder speed settings of the monitor to revert to factory default may lead to misinterpretation of the fetal trace by the user and to potential incorrect clinical decisions
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- Philips Healthcare issued an "Urgent-Medical Device Correction" notification dated May 15, 2009 to Customers in the United States via UPS and notifications outside the US will be managed by Philips representatives in each affected geography. Customers were asked to follow the guidelines in the Action to be Taken by Customer/User section of the Correction Notice until they receive their software upgrade. Questions can be made to Philips response center at 1-800-722-9377.
- Quantity in commerce
- 4,788 units
- Distribution
- Worldwide Distribution
- Product codes and lots
- Units with serial numbers in the range of DE53001002 through DE53107383.