Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1585-2009

Ansell, Derma Prene Ultra, Powder-Free, Smooth Surgical Gloves, REF 8514, Size 7, 4 boxes of 50, Sterile...

Ansell Healthcare Products LLC / initiated 2009-05-01 / Terminated / root cause: manufacturing process

Ansell Healthcare Products LLC began a recall of Ansell, Derma Prene Ultra, Powder-Free, Smooth Surgical Gloves, REF 8514, Size 7, 4 boxes of 50, Sterile, Distributed by Ansell Healthcare, 1635 Industrial... on . The reason FDA records is "Premature degradation which could result in tearing during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1585-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ansell Healthcare Products LLC
Product
Ansell, Derma Prene Ultra, Powder-Free, Smooth Surgical Gloves, REF 8514, Size 7, 4 boxes of 50, Sterile, Distributed by Ansell Healthcare, 1635 Industrial Road, Dothan, AL, 36303, USA, Made in Sri Lanka. The device is made of synthetic rubber and is intended to be worn by operating room personnel to protect a surgical wound from contamination and for use in handling chemotherapy drugs.
Product code
KGO Surgeon'S Gloves
Reason for recall
Premature degradation which could result in tearing during use.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated the recall on May 1, 2009 by letter to Health Canada and by letter dated May 4, 2009 to its US and Canadian consignees. The letters described the affected product, issue and actions for customers. The recall was extended to the hospital and clinic level requesting that the product be returned. Direct questions to your Ansell Healthcare Products LLC representative.
Quantity in commerce
560,000 total units--all lots.
Distribution
Nationwide Distribution and Canada.
Product codes and lots
Lot Numbers: 0810003305, 0180007305, 0810007805, 0810015505 and 0810006705.